Tuesday, August 24, 2010

The HPLC Stereoselective Resolution of N-2,4-Dinitrophenylated (DNP) Amine-Containing Enantiomers on Teicoplanin Bonded Phase Using the Methanol-Based Mobile Phase

Abstract

A variety of compounds containing an amine group (i.e., amino acids and amino alcohols) are fast and much better than baseline enantioresolved on a teicoplanin bonded chiral phase using the methanol-based mobile phase after their pre-column derivatization with 2,4-dinitrofluorobenzene (DNFB) in alkaline medium. The resolution is found relatively insensitive to the structural variations of the analyte and is enhanced as the stereogenic center of analyte is close to the nitro group substituted aromatic moiety of the tagging reagent for stronger pgr-pgr interaction. The resolution is either not observed or unsatisfactory using the acetonitrile-based mobile phase.
Keywords  Column liquid chromatography Enantioresolution Amino acids 2 - 4-Dinitrofluorobenzene Teicoplanin

Fulltext Preview

Image of the first page of the fulltext document

Validation of a Reversed-Phase HPLC Method for Quantitative Amino Acid Analysis

Milano Research Centre, Bracco Imaging SpA, Milan, Italy
Dr. Federico Maisano, CRM/pharmatox, Bracco imaging SPA, Via Folli, 50, 20134 Milan, Italy (phone +39 02 21772369; fax +39 02 21772794; email: federico.maisano@bracco.com).
A semi-automated method for amino acid derivatization and analysis has been validated for use in analysis of protein biopharmaceuticals. The method includes protein hydrolysis, o-phthalaldehyde derivatization, and reversed-phase high-performance liquid chromatography analysis in a general-purpose UV-visible high-performance liquid chromatography system. Amino-acid derivatization is performed automatically by the high-performance liquid chromatography autosampler right before injection. The required validation parameters, i.e., specificity, linearity, accuracy, precision, limit of detection, and limit of quantification, were studied for bovine serum albumin and for a recombinant human Fab fragment. The method can be employed as an absolute quantification method for determination of extinction coefficients of recombinant proteins.
Keywords: Amino acid analysis, OPA-derivatization, reverse-phase HPLC, validation
Amino-acid analysis has a long history in the characterization of protein-based products, since it provides information on the product concentration without referring to an external protein standard and it is independent from the shape and the charge of the protein. In addition, the determined amino-acid composition can confirm sample identity and gives a measure of sample purity. Furthermore, when combined with absorbance measurements, it allows the determination of extinction coefficients under various conditions.1 For protein conjugates, where the synthetic counterpart modifies the protein absorption properties, amino-acid analysis may be required as the only reliable quantification method.
However, in spite of these features, few laboratories can perform such analysis in a reliable and quantitative way, due to the need for specialized equipment and skills. Usually, techniques based on ion-exchange separation coupled with post-column derivatization (e.g., with nin-hydrin, the “classical” method) are considered more precise1 than those based on pre-column derivatization and reversed-phase high-performance liquid chromatography (RP-HPLC), because the latter techniques imply extensive sample manipulation before analysis and are affected by the limited stability of the preformed derivatives.2 However, such RP-HPLC-based methods have the advantage of being accessible to most analytical laboratories, since they do not require expensive dedicated instruments. In addition, manufacturing of dedicated instruments is being halted, making the availability of validated pre-column methods even more important.
In this paper, we describe the validation of a method that takes advantage of robotic sample derivatization, thereby limiting considerably the manual manipulation of samples. Another advantage of automation is that derivatization is performed just before the injection; therefore, the time from reaction to injection is kept absolutely constant for all samples, thus avoiding differential degradation of labile derivatives. We have studied the performance characteristics in terms of specificity, linearity, accuracy, precision, limit of detection, and limit of quantification for bovine serum albumin (BSA) and for a recombinant human Fab (rFab) fragment, whose extinction coefficient needs to be determined.
Protein samples were hydrolyzed, then automatically derivatized with o-phthalaldehyde (OPA) and in-line analyzed by RP-HPLC with ultraviolet-visible (UV-Vis) detection, according to a method published in an Agilent application note.3
Reagents, Solvents, and Materials
Sodium phosphate monobasic monohydrate, sodium hydroxide, boric acid, acetonitrile (LC grade), and methanol (LC grade) were obtained from Merck KGaA (Darmstadt, Germany). OPA reagent was prepared as described (Agilent art. 5061-3335, Palo Alto, CA). Borate buffer was prepared by adjusting 0.4 N boric acid to pH 10.2 with NaOH. Constant-boiling HCl was obtained from Sigma-Aldrich (St. Louis, MO). Chromatographic-grade water was produced by a Milli-Q system (Millipore, Billerica, MA)
Disposable glass test tubes (50 × 6 mm) and hydrolysis reaction vials (25 × 120 mm) with Mininert valves were from Kimble Glass, Inc., and Kontes Glass Co. (Vineland, NJ). Amber wide-opening vials, glass conical inserts with polymer feet, and screw caps were from Agilent.
Albumin standard solution (2 mg/mL) was supplied by Pierce Biotechnology (Rockford, IL), while amino acid standard mixtures at the concentration of 1 nmol/μL and 250 pmol/μL were from Agilent. The internal standard l-norvaline was obtained from Sigma-Aldrich. A recombinant Fab fragment (rFab) was obtained from the research laboratories of Bracco Imaging (Milan, Italy).
Amino Acid Standard Solutions
Amino acid standard samples were prepared by mixing 95 μL of the 250 pmol/μL amino acid standard mixture with 5 μL of 10 mM norvaline and analyzed directly by RP-HPLC, within 24 h from preparation. Solutions for linearity study were prepared in duplicate by diluting the 1 nmol/μL amino acid standard solution, and contained 20, 50, 130, 250, or 500 pmol/μL of amino acid standard mixture together with 0.5 mM norvaline.
Protein Samples
Glass test tubes (50 × 6 mm) were marked with incisions and soaked in a detergent solution for at least 12 h. They were rinsed thoroughly in Milli-Q water and dried in an oven at 80°C. Protein samples (7–75 μg) were transferred into the glass test tubes and spiked with 0.5 mM norvaline. They were quickly spun in a low-velocity centrifuge, then frozen and dried in a lyophilizer. Samples were then transferred into the reaction vial containing 0.5 mL of constant-boiling HCl on the bottom. Up to 12 test tubes could be accommodated in a reaction vial. The reaction vial was tightly closed and transferred into a pre-heated oven at 110°C for 18 h. The reaction vial was cooled at room temperature, then carefully opened under an aspirated hood. The test tubes were centrifuged and dried again in the lyophilizer to remove any liquid traces (condensed vapors). The dried residues were dissolved in 100 μL of 0.1 N HCl and transferred into the HPLC glass insert vials.
Instrumentation
Analyses were performed using an Agilent 1100 Liquid Chromatograph, equipped with a binary pump delivery system (G1312A), robotic autosampler (G1313A), column thermostat (G1316A) and multi-wavelength detector (G1365A).
Analytical Procedure
Chromatography conditions were in accordance with the Agilent method.2 Briefly, the hydrolyzed samples and the norvaline-spiked amino acid standard solutions were automatically derivatized with OPA by programming the robotic autosampler (Table 1). After derivatization, an amount equivalent to 2.5 μL of each sample was injected on a Zorbax Eclipse-AAA column, 5 μm, 150 × 4.6 mm (Agilent), at 40°C, with detection at λ = 338 nm. Mobile phase A was 40 mM NaH2PO4, adjusted to pH 7.8 with NaOH, while mobile phase B was acetonitrile/methanol/ water (45/45/10 v/v/v). The separation was obtained at a flow rate of 2 mL/min with a gradient program that allowed for 1.9 min at 0% B followed by a 16.3-min step that raised eluent B to 53%. Then washing at 100% B and equilibration at 0% B was performed in a total analysis time of 26 min.
TABLE 1
TABLE 1
Autosampler Programming Instructions
Acid hydrolysis is a crucial step that considerably influences amino-acid recovery. In fact, during acid hydrolysis, tryptophan and cysteine are destroyed and serine and threonine are also partially lost, while methionine can undergo oxidation. Moreover, some amino acids such as glycine and serine are common contaminants; therefore, their quantification needs careful subtraction of average responses in blank runs, which, in the case of glycine, is also complicated by the fact that this residue is known to give rise to multiple derivatives after OPA reaction.2 Therefore, the validation parameters were estimated using the following seven best-recovered amino acids: Asx (Asn+Asp), Glx (Glu+Gln), Arg, Ala, Phe, Leu, and Lys.4
In order to fully assess the method’s performance, both a standard amino acid mixture and a reference protein (e.g., BSA) should be assayed along with the product. The standard amino acid mixture (Figure 11)) enables the verification of the HPLC method’s performance, including derivatization, while the reference protein samples(Figure 22)) assess the completeness of the hydrolysis step. In addition, L-norvaline, which is added as the internal standard, provides a control for sample-to-sample variability.
FIGURE 1
FIGURE 1
Example of a standard amino acid mixture analysis at a concentration of 250 pmol/μl.
FIGURE 2
FIGURE 2
Example of a protein hydrolysate analysis for one of the 35-μg BsA samples.
Specificity
Specificity was documented by comparing retention times obtained in the standard amino acid mixture (five samples) with those obtained from the reference protein samples (three samples). Results are reported in Table 2. The minimal difference between retention times (<0.1%) allows confident, highly specific, peak identification. Usually, a difference within ±3% is considered acceptable between retention times of the same amino acids present in the standard mixture and in the hydrolyzed sample.5
TABLE 2
TABLE 2
Specificity Evaluation Comparing Retention Times
Linearity and Range
Linearity was studied in the range from 20 to 500 pmol/μL of standard amino acids and from 1.5 to 15 pmol/μL of rFab, corresponding approximately to 20–200% of the test concentration. Five concentration points were assayed in duplicate. Both the standard amino acid mixture and the test product showed good linearity in the tested range. The area response obeyed the equation y = mx + C, where the intercept C was zero within 95% confidence limits and the square correlation coefficient (R2) was always greater then 0.985 (Table 3).
TABLE 3
TABLE 3
Linearity Data for Amino Acid Standard Mixture and rFab
Accuracy
Accuracy is defined as the agreement between the found value and the true, independently determined, concentration value. It was studied on BSA samples prepared from a calibrated standard, which is normally used in protein determination assays. The accuracy of the method was evaluated considering two different parameters: the absolute percent error and the recovery. The absolute percent error (%ABS) was calculated for each well-recovered amino acid. It was always less then 7%, and the average percent error was less then 2.5% (Table 4). Based on the found amino acid concentrations and the known composition of BSA, the percent recovery of BSA was calculated. Considering each single amino acid, the recovery was found within the range 97–108%, while the average recovery, considering all the amino acids, was 102% (Table 5), which is largely within the 90–110% range that is considered acceptable.5
TABLE 4
TABLE 4
Absolute Percent Error (%ABS) of Found to Theoretical (Nt) Residue Numbers for the BSA Molecule
TABLE 5
TABLE 5
BSA Recovery, as Calculated from Single Amino Acids
Precision
Precision was measured as repeatability and intermediate precision. Reproducibility, which refers to the use of the analytical procedure in different laboratories, was beyond the scope of the present study.
Repeatability was studied on six injections (and derivatization) of the same BSA sample. The mean accuracy of each well-recovered amino acid was calculated. The percentage coefficient of variation (%CV) for each well-recovered amino acid was less than 2.5% (Table 6), compared with an acceptance range of 5%.
TABLE 6
TABLE 6
Repeatability (Numbers of Residues)
Intermediate precision was studied by running the whole method on three different days. Each day, three equivalent BSA samples were prepared, hydrolyzed, derivatized, and injected. The %CV for each well-recovered amino acid was generally not greater than 2.5%, with only one exception for arginine at 8%, as reported in Table 7.
TABLE 7
TABLE 7
Intermediate Precision of Well-Recovered Amino Acid Residues (Numbers of Residues)
Quantification and Detection Limits
The limit of quantification (LOQ) and the limit of detection (LOD) were calculated on the basis of the standard deviation of the response and the slope obtained from the linearity plot of each well-recovered amino acid of the standard mixture, as described in the relevant ICH guideline.6 LOQ and LOD were calculated as 3.3α/S and 10α/S, respectively, where α is the standard deviation of the y-intercept and S is the slope of regression line. The calculated values of LOQ and LOD for each considered residue are reported in Table 8.
TABLE 8
TABLE 8
LOQ and LOD Values (pmol/μL)
Method specificity was demonstrated by comparing the retention times for the considered amino acids in hydrolyzed BSA samples and amino acid standard solutions. Given the excellent reproducibility of retention times between amino acid standard solution and protein samples, the method proved highly specific.
Linearity was demonstrated for single amino acid response both in recombinant Fab and in standard amino acid samples in a range of 25–150% of the usually analyzed amount. The square correlation coefficient was above 0.985, with only one case (arginine) below 0.99.
The LOQ and the LOD were in the range 33–60 pmol/μL and 11–20 pmol/μL, respectively, depending on the amino acid under consideration.
Accuracy was determined by comparison of found amino acid residue numbers with those obtained from BSA known composition. The average absolute error was 2.33%. In addition, BSA recovery was determined both from single amino acids (range 97–108%) and as an average value (102%).
Repeatability was measured by carrying out analyses on six injections (including derivatization) of the same hydrolyzed BSA sample. The %CV for each considered amino acid was less than 5%.
Intermediate precision was determined on three different days, by processing three equivalent BSA samples each day. For single amino acid recovery, the %CV for each considered amino acid was less than 8%, with only one case (arginine) above 2.5%.
The determined validation parameters are in the commonly acceptable ranges for this kind of analysis, and allow the use of the method with various recombinant proteins (e.g., antibody fragments) both as an identity test and for extinction coefficient determination when the sequence in known.
An important contribution to these results came from the automation adopted for amino acid derivatization, which guarantees highly reproducible reaction times and lack of degradation. In addition, the short reaction times seem not to give rise to multiple derivatives, as previously observed.2 Also the use of absorbance instead of fluorescence detection did not negatively influence the performance characteristics of the method.
1. Anders JC. Advances in amino acid analysis. BioPharm Int 2002;4:32–39.
2. Mengerink Y, Kutlán D, Tóth F, Csámpai A, Molnár-Perl I. Advances in the evaluation of the stability and characteristics of the amino acid and amine derivatives obtained with the o-phthaldialdehyde/3-mercaptopropionic acid and o-phthaldialdehyde/N-acetyl-l-cysteine reagents: High-performance liquid chromatography-mass spectrometry study. J Chromatogr A 2002;949:99–124. [PubMed]
3. Henderson JW, Ricker RD, Bidlingmeyer BA, Woodward C. Agilent Technical Note 1999;5980–1193E.
4. Anders JC, Parten BF, Petrie GE, Marlowe RL, McEntire JE. Using amino acid analysis to determine absorptivity constants: A validation case study using bovine serum albumin. BioPharm Int 2003;2:30–37.
5. Reason AJ. Validation of amino acid analysis methods. In Smith BJ (ed), Protein Sequencing Protocols, 2nd Edition: Methods in Molecular Biolog y, Vol. 211. Totowa, NJ: Humana Press, 2003:181–194.
6. International Conference of Harmonisation (ICH) Harmonised Tripartite Guideline—Validation of Analytical Procedures: Methodology Q2B. November 1996.

Monday, August 23, 2010

HPLC Nomenclature

*Indicates a definition adapted from: L.S. Ettre, Nomenclature for Chromatography, Pure Appl. Chem. 65: 819-872 [1993], © 1993 IUPAC; an updated version of this comprehensive report is available in the Orange Book, Chapter 9: Separations [1997] at: <http://www.iupac.org/publications/analytical_compendium>.
Alumina
A porous, particulate form of aluminum oxide [Al203] used as a stationary phase in normal-phase adsorption chromatography. Alumina has a highly active basic surface; the pH of a 10% aqueous slurry is about 10. It is successively washed with strong acid to make neutral and acidic grades [slurry pH 7.5 and 4, resp.]. Alumina is more hygroscopic than silica. Its activity is measured according to the Brockmann† scale for water content; e.g., Activity Grade I contains 1% H2O.
†H. Brockmann and H. Schodder, Ber. 74: 73 (1941).
Baseline*
The portion of the chromatogram recording the detector response when only the mobile phase emerges from the column.
Cartridge
A type of column, without endfittings, that consists simply of an open tube wherein the packing material is retained by a frit at either end. SPE cartridges may be operated in parallel on a vacuum-manifold. HPLC cartridges are placed into a cartridge holder that has fluid connections built into each end. Cartridge columns are easy to change, less expensive, and more convenient than conventional columns with integral endfittings.
Chromatogram*
A graphical or other presentation of detector response or other quantity used as a measure of the concentration of the analyte in the effluent versus effluent volume or time. In planar chromatography [e.g., thin-layer chromatography or paper chromatography], chromatogram may refer to the paper or layer containing the separated zones.
Chromatography*
A dynamic physicochemical method of separation in which the components to be separated are distributed between two phases, one of which is stationary [the stationary phase] while the other [the mobile phase] moves relative to the stationary phase.
Column Volume* [Vc]
The geometric volume of the part of the tube that contains the packing [internal cross-sectional area of the tube multiplied by the packed bed length, L]. The interparticle volume of the column, also called the interstitial volume, is the volume occupied by the mobile phase between the particles in the packed bed. The void volume [V0] is the total volume occupied by the mobile phase, i.e. the sum of the interstitial volume and the intraparticle volume [also called pore volume].
Detector* [see Sensitivity]
A device that indicates a change in the composition of the eluent by measuring physical or chemical properties [e.g., UV/visible light absorbance, differential refractive index, fluorescence, or conductivity]. If the detector’s response is linear with respect to sample concentration, then, by suitable calibration with standards, the amount of a component may be quantitated. Often, it may be beneficial to use two different types of detectors in series. In this way, more corroboratory or specific information may be obtained about the sample analytes. Some detectors [e.g., electrochemical, mass spectrometric] are destructive; i.e., they effect a chemical change in the sample components. If a detector of this type is paired with a non-destructive detector, it is usually placed second in the flow path.
Display
A device that records the electrical response of a detector on a computer screen in the form of a chromatogram. Advanced data recording systems also perform calculations using sophisticated algorithms, e.g., to integrate peak areas, subtract baselines, match spectra, quantitate components, and identify unknowns by comparison to standard libraries.
Efficiency [H, see Plate Number, Resolution, Sensitivity, Speed]
A measure of a column’s ability to resist the dispersion of a sample band as it passes through the packed bed. An efficient column minimizes band dispersion or bandspreading. Higher efficiency is important for effective separation, greater sensitivity, and/or identification of similar components in a complex sample mixture.
Nobelists Martin and Synge, by analogy to distillation, introduced the concept of plate height [H, or H.E.T.P., height equivalent to a theoretical plate] as a measure of chromatographic efficiency and as a means to compare column performance.† Presaging HPLC and UPLC technology, they recognized that a homogeneous bed packed with the smallest possible particle size [requiring higher pressure] was key to maximum efficiency. The relation between column and separation system parameters that affect bandspreading
was later described in an equation by van Deemter.††
Chromatographers often refer to a quantity that they can calculate easily and directly from measurements made on a chromatogram, namely plate number [N], as efficiency. Plate height is then determined from the ratio of the length of the column bed to N [H = L/N; methods of calculating N from a chromatogram are shown in Figure U]. It is important to note that calculation of N or H using these methods is correct only for isocratic conditions and cannot be used for gradient separations.
†A.J.P. Martin and R.M. Synge, Biochem. J. 35: 1358-1368 [1941]
††J.J. van Deemter, F. J. Zuiderweg and A. Klinkenberg, Chem. Eng. Sci. 5: 271-289 [1956]
Eluate
The portion of the eluent that emerges from the column outlet containing analytes in solution. In analytical HPLC, the eluate is examined by the detector for the concentration or mass of analytes therein. In preparative HPLC, the eluate is collected continuously in aliquots at uniform time or volume intervals, or discontinuously only when a detector indicates the presence of a peak of interest. These fractions are subsequently processed to obtain purified compounds.
Eluent
The mobile phase [see Elution Chromatography].
Eluotropic Series
A list of solvents ordered by elution strength with reference to specified analytes on a standard sorbent. Such a series is useful when developing both isocratic and gradient elution methods. Trappe coined this term after showing that a sequence of solvents of increasing polarity could separate lipid fractions on alumina.† Later, Snyder measured and tabulated solvent strength parameters for a large list of solvents on several normal-phase LC sorbents.†† Neher created a very useful nomogram by which equi-eluotropic
[constant elution strength] mixtures of normal-phase solvents could be chosen to optimize the selectivity of TLC separations.†††
A typical normal-phase eluotropic series would start at the weak end with non-polar aliphatic hydrocarbons, e.g., pentane or hexane, then progress successively to benzene [an aromatic hydrocarbon], dichloromethane [a chlorinated hydrocarbon], diethyl ether, ethyl acetate [an ester], acetone [a ketone], and, finally, methanol [an alcohol] at the strong end [see Figure R-1].
†W. Trappe, Biochem. Z. 305: 150 [1940]
††L. R. Snyder, Principles of Adsorption Chromatography, Marcel Dekker [1968], pp. 192-197
†††R. Neher in G.B. Marini-Bettòlo, ed., Thin-Layer Chromatography, Elsevier [1964] pp. 75-86.
Elute* [verb]
To chromatograph by elution chromatography. The process of elution may be stopped while all the sample components are still on the chromatographic bed [planar thin-layer or paper chromatography] or continued until the components have left the chromatographic bed [column chromatography].
Note: The term elute is preferred to develop [a term used in planar chromatography], to avoid confusion with the practice of method development, whereby a separation system [the combination of mobile and stationary phases] is optimized for a particular separation.
Elution Chromatography*
A procedure for chromatographic separation in which the mobile phase is continuously passed through the chromatographic bed. In HPLC, once the detector baseline has stabilized and the separation system has reached equilibrium, a finite slug of sample is introduced into the flowing mobile phase stream. Elution continues until all analytes of interest have passed through the detector.
Elution Strength
A measure of the affinity of a solvent relative to that of the analyte for the stationary phase. A weak solvent cannot displace the analyte, causing it to be strongly retained on the stationary phase. A strong solvent may totally displace all the analyte molecules and carry them through the column unretained. To achieve a proper balance of effective separation and reasonable elution volume, solvents are often blended to set up an appropriate competition between the phases, thereby optimizing both selectivity and
separation time for a given set of analytes [see Selectivity].
Dipole moment, dielectric constant, hydrogen bonding, molecular size and shape, and surface tension may give some indication of elution strength. Elution strength is also determined by the separation mode. An eluotropic series of solvents may be ordered by increasing strength in one direction under adsorption or normal-phase conditions; that order may be nearly opposite under reversed-phase partition conditions [see Figure R-1].
Fluorescence Detector
Fluorescence detectors excite a sample with a specified wavelength of light. This causes certain compounds to fluoresce and emit light at a higher wavelength. A sensor, set to a specific emission wavelength and masked so as not to be blinded by the excitation source, collects only the emitted light. Often analytes that do not natively fluoresce may be derivatized to take advantage of the high sensitivity and selectivity of this form of detection, e.g., AccQ•Tag™ derivatization of amino acids.
Flow Rate*
The volume of mobile phase passing through the column in unit time. In HPLC systems, the flow rate is set by the controller for the solvent delivery system [pump]. Flow rate accuracy can be checked by timed collection and measurement of the effluent at the column outlet. Since a solvent’s density varies with temperature, any calibration or flow rate measurement must take this variable into account. Most accurate determinations are made, when possible, by weight, not volume.
Uniformity [precision] and reproducibility of flow rate is important to many LC techniques, especially in separations where retention times are key to analyte identification, or in gel-permeation chromatography where calibration and correlation of retention times are critical to accurate molecular-weight-distribution measurements of polymers.
Often, separation conditions are compared by means of linear velocity, not flow rate. The linear velocity is calculated by dividing the flow rate by the cross-sectional area of the column. While flow rate is expressed in volume/time [e.g., mL/min], linear velocity is measured in length/time [e.g., mm/sec].
Gel-Permeation Chromatography*
Separation based mainly upon exclusion effects due to differences in molecular size and/or shape. Gelpermeation chromatography and gel filtration chromatography describe the process when the stationary phase is a swollen gel. Both are forms of size-exclusion chromatography. Porath and Flodin first described gel-filtration using dextran gels and aqueous mobile phases for the size-based separation of biomolecules.† Moore applied similar principles to the separation of organic polymers by size in solution using
organic-solvent mobile phases on porous polystyrene-divinylbenzene polymer gels.††
†J. Porath, P. Flodin, Nature 183: 1657-1659 [1959]
††J.C. Moore, U.S. Patent 3,326,875 [filed Jan. 1963; issued June 1967]
Gradient
The change over time in the relative concentrations of two [or more] miscible solvent components that form a mobile phase of increasing elution strength. A step gradient is typically used in solid-phase extraction; in each step, the eluent composition is changed abruptly from a weaker mobile phase to a stronger mobile phase. It is even possible, by drying the SPE sorbent bed in between steps, to change from one solvent to another immiscible solvent.
A continuous gradient is typically generated by a low- or high-pressure mixing system [see Figures J-2 and J-3] according to a pre-determined curve [linear or non-linear] representing the concentration of the stronger solvent B in the initial solvent A over a fixed time period. A hold at a fixed isocratic solvent composition can be programmed at any time point within a continuous gradient. At the end of a separation, the gradient program can also be set to return to the initial mobile phase composition to re-equilibrate the column in preparation for the injection of the next sample. Sophisticated HPLC systems can blend as many as four or more solvents [or solvent mixtures] into a continuous gradient.
Injector [Autosampler, Sample Manager]
A mechanism for accurately and precisely introducing [injecting] a discrete, predetermined volume of a sample solution into the flowing mobile phase stream. The injector can be a simple manual device, or a sophisticated autosampler that can be programmed for unattended injections of many samples from an array of individual vials or wells in a predetermined sequence. Sample compartments in these systems may even be temperature controlled to maintain sample integrity over many hours of operation.
Most modern injectors incorporate some form of syringe-filled sample loop that can be switched on- or offline by means of a multi-port valve. A well-designed, minimal-internal-volume injection system is situated as close to the column inlet as possible and minimizes the spreading of the sample band. Between sample injections, it is also capable of being flushed to waste by mobile phase, or a wash solvent, to prevent carryover [contamination of the present sample by a previous one].
Samples are best prepared for injection, if possible, by dissolving them in the mobile phase into which they will be injected; this may prevent issues with separation and/or detection. If another solvent must be used, it is desirable that its elution strength be equal to or less than that of the mobile phase. It is often wise to mix a bit of a sample solution with the mobile phase offline to test for precipitation or miscibility issues that might compromise a successful separation.
Inlet
The end of the column bed where the mobile phase stream and sample enter. A porous, inert frit retains the packing material and protects the sorbent bed inlet from particulate contamination. Good HPLC practice dictates that samples and mobile phases should be particulate-free; this becomes imperative for small-particle columns whose inlets are much more easily plugged. If the column bed inlet becomes clogged and exhibits higher-than-normal backpressure, sometimes, reversing the flow direction while directing the effluent to waste may dislodge and flush out sample debris that sits atop the frit. If the
debris has penetrated the frit and is lodged in the inlet end of the bed itself, then the column has most likely reached the end of its useful life.
Ion-Exchange Chromatography* [see section: Separations Based on Charge]
This separation mode is based mainly on differences in the ion-exchange affinities of the sample components. Separation of primarily inorganic ionic species in water or buffered aqueous mobile phases on small particle, superficially porous, high-efficiency, ion-exchange columns followed by conductometric or electrochemical detection is referred to as ion chromatography [IC].
Isocratic Elution*
A procedure in which the composition of the mobile phase remains constant during the elution process.
Liquid Chromatography* [LC]
A separation technique in which the mobile phase is a liquid. Liquid chromatography can be carried out either in a column or on a plane [TLC or paper chromatography]. Modern liquid chromatography utilizing smaller particles and higher inlet pressure was termed high-performance (or high-pressure) liquid chromatography [HPLC] in 1970. In 2004, ultra-performance liquid chromatography dramatically raised the performance of LC to a new plateau [see UPLC Technology].
Mobile Phase* [see Eluate, Eluent]
A fluid that percolates, in a definite direction, through the length of the stationary-phase sorbent bed. The mobile phase may be a liquid [liquid chromatography] or a gas [gas chromatography] or a supercritical fluid [supercritical-fluid chromatography]. In gas chromatography the expression carrier gas may be used for the mobile phase. In elution chromatography, the mobile phase may also be called the eluent, while the word eluate is defined as the portion of the mobile phase that has passed through the sorbent bed and contains the compounds of interest in solution.
Normal-Phase Chromatography*
An elution procedure in which the stationary phase is more polar than the mobile phase. This term is used in liquid chromatography to emphasize the contrast to reversed-phase chromatography.
Peak* [see Plate Number]
The portion of a differential chromatogram recording the detector response while a single component is eluted from the column. If separation is incomplete, two or more components may be eluted as one unresolved peak. Peaks eluted under optimal conditions from a well-packed, efficient column, operated in a system that minimizes bandspreading, approach the shape of a Gaussian distribution. Quantitation is usually done by measuring the peak area [enclosed by the baseline and the peak curve]. Less often, peak height [the distance measured from the peak apex to the baseline] may be used for quantitation. This procedure requires that both the peak width and the peak shape remain constant.
Plate Number* [N, see Efficiency]
A number indicative of column performance [mechanical separation power or efficiency, also called plate count, number of theoretical plates, or theoretical plate number]. It relates the magnitude of a peak’s retention to its width [variance or bandspread]. In order to calculate a plate count, it is assumed that a peak can be represented by a Gaussian distribution [a statistical bell curve]. At the inflection points [60.7% of peak height], the width of a Gaussian curve is twice the standard deviation [σ] about its mean [located at the peak apex]. As shown in Figure U, a Gaussian curve’s peak width measured at other fractions of peak height can be expressed in precisely defined multiples of σ. Peak retention [retention volume, VR, or retention time, tR] and peak width must be expressed in the same units, because N is a dimensionless number. Note that the 5 sigma method of calculating N is a more stringent measure of column homogeneity and performance, as it is more severely affected by peak asymmetry. Computer data stations can automatically delineate each resolved peak and calculate its corresponding plate number.
Preparative Chromatography
The process of using liquid chromatography to isolate a compound in a quantity and at a purity level sufficient for further experiments or uses. For pharmaceutical or biotechnological purification processes, columns several feet in diameter can be used for multiple kilograms of material. For isolating just a few micrograms of a valuable natural product, an analytical HPLC column is sufficient. Both are preparative chromatographic approaches, differing only in scale [see section on HPLC Scale and Table A].
Resolution* [Rs, see Selectivity]
The separation of two peaks, expressed as the difference in their corresponding retention times, divided by their average peak width at the baseline. Rs = 1.25 indicates that two peaks of equal width are just separated at the baseline. When Rs = 0.6, the only visual indication of the presence of two peaks on a chromatogram is a small notch near the peak apex. Higher efficiency columns produce narrower peaks and improve resolution for difficult separations; however, resolution increases by only the square root of N. The most powerful method of increasing resolution is to increase selectivity by altering the mobile/stationary phase combination used for the chromatographic separation [see section on Chemical Separation Power].
Retention Factor* [k]
A measure of the time the sample component resides in the stationary phase relative to the time it resides in the mobile phase; it expresses how much longer a sample component is retarded by the stationary phase than it would take to travel through the column with the velocity of the mobile phase. Mathematically, it is the ratio of the adjusted retention time [volume] and the hold-up time [volume]: k = tR'/tM [see Retention Time and Selectivity].
Note: In the past, this term has also been expressed as partition ratio, capacity ratio, capacity factor, or mass distribution ratio and symbolized by k'.
Retention Time* [tR]
The time between the start of elution [typically, in HPLC, the moment of injection or sample introduction] and the emergence of the peak maximum. The adjusted retention time, tR', is calculated by subtracting from tR the hold-up time [tM, the time from injection to the elution of the peak maximum of a totally unretained analyte].
Reversed-Phase Chromatography*
An elution procedure used in liquid chromatography in which the mobile phase is significantly more polar than the stationary phase, e.g. a microporous silica-based material with alkyl chains chemically bonded to its accessible surface. Note: Avoid the incorrect term reverse phase. [See Reference 4 for some novel ideas on the mechanism of reversed-phase separations.]
Selectivity [Separation Factor, σ]
A term used to describe the magnitude of the difference between the relative thermodynamic affinities of a pair of analytes for the specified mobile and stationary phases that comprise the separation system. The proper term is separation factor [σ]. It equals the ratio of retention factors, k2/k1 [see Retention Factor]; by definition, σ is always ≥ 1. If σ = 1, then both peaks co-elute, and no separation is obtained. It is important in preparative chromatography to maximize α for highest sample loadability and throughput. [see section on Chemical Separation Power]
Sensitivity* [S]
The signal output per unit concentration or unit mass of a substance in the mobile phase entering the detector, e.g., the slope of a linear calibration curve [see Detector]. For concentration-sensitive detectors [e.g., UV/VIS absorbance], sensitivity is the ratio of peak height to analyte concentration in the peak. For mass-flow-sensitive detectors, it is the ratio of peak height to unit mass. If sensitivity is to be a unique performance characteristic, it must depend only on the chemical measurement process, not upon scale factors.
The ability to detect [qualify] or measure [quantify] an analyte is governed by many instrumental and chemical factors. Well-resolved peaks [maximum selectivity] eluting from high efficiency columns [narrow peak width with good symmetry for maximum peak height] as well as good detector sensitivity and specificity are ideal. Both the separation system interference and electronic component noise should also be minimized to achieve maximum sensitivity.
Solid-Phase Extraction [SPE]
A sample preparation technique that uses LC principles to isolate, enrich, and/or purify analytes from a complex matrix applied to a miniature chromatographic bed. Offline SPE is done [manually or via automation] with larger particles in individual plastic cartridges or in micro-elution plate wells, using low positive pressure or vacuum to assist flow. Online SPE is done with smaller particles in miniature HPLC columns using higher pressures and a valve to switch the SPE column online with the primary HPLC column, or offline to waste, as appropriate.
SPE methods use step gradients [see Gradient] to accomplish bed conditioning, sample loading, washing, and elution steps. Samples are loaded typically under conditions where the k of important analytes is as high as possible, so that they are fully retained during loading and washing steps. Elution is then done by switching to a much stronger solvent mixture [see Elution Strength]. The goal is to remove matrix interferences and to isolate the analyte in a solution, and at a concentration, suitable for subsequent analysis.
Speed [see Efficiency, Flow Rate, Resolution]
A benefit of operating LC separations at higher linear velocities using smaller-volume, smaller-particle analytical columns, or larger-volume, larger-particle preparative columns. Order-of-magnitude advances in LC speed came in 1972 [with the use of 10 μm particles and pumps capable of delivering accurate mobile-phase flow at 6000 psi], in 1976 [with 75-μm preparative columns operated at a flow rate of 500 mL/min], and in 2004 [with the introduction of UPLC technology—1.7 μm-particle columns operated at 15,000 psi].†
High-speed analytical LC systems must not only accommodate higher pressures throughout the fluidics; injector cycle time must be short; gradient mixers must be capable of rapid turnaround between samples; detector sensors must rapidly respond to tiny changes in eluate composition; and data systems must collect the dozens of points each second required to plot and to quantitate narrow peaks accurately.
Together, higher resolution, higher speed, and higher efficiency typically deliver higher throughput. More samples can be analyzed in a workday. Larger quantities of compound can be purified per run or per process period.
†See #3 on list of References for Further Reading above.
Stationary Phase*
One of the two phases forming a chromatographic system. It may be a solid, a gel, or a liquid. If a liquid, it may be distributed on a solid. This solid may or may not contribute to the separation process. The liquid may also be chemically bonded to the solid [bonded phase] or immobilized onto it [immobilized phase].
The expression chromatographic bed or sorbent may be used as a general term to denote any of the different forms in which the stationary phase is used.
The use of the term liquid phase to denote the mobile phase in LC is discouraged. This avoids confusion with gas chromatography where the stationary phase is called a liquid phase [most often a liquid coated on a solid support].
Open-column liquid-liquid partition chromatography [LLC] did not translate well to HPLC. It was supplanted by the use of bonded-phase packings. LLC proved incompatible with modern detectors because of problems with bleed of the stationary-phase-liquid coating off its solid support, thereby contaminating the immiscible liquid mobile phase.
UPLC® Technology
The use of a high-efficiency LC system holistically designed to accommodate sub-2 μm particles and very high operating pressure is termed ultra-performance liquid chromatography [UPLC technology].† The major benefits of this technology are significant improvements in resolution over HPLC, and/or faster run times while maintaining the resolution seen in an existing HPLC separation.

HPLC Separation Modes

In general, three primary characteristics of chemical compounds can be used to create HPLC separations. They are:
• Polarity
• Electrical Charge
• Molecular Size
First, let’s consider polarity and the two primary separation modes that exploit this characteristic: normal phase and reversed-phase chromatography.
Separations Based on Polarity
A molecule’s structure, activity, and physicochemical characteristics are determined by the arrangement of its constituent atoms and the bonds between them. Within a molecule, a specific arrangement of certain atoms that is responsible for special properties and predictable chemical reactions is called a functional group. This structure often determines whether the molecule is polar or non-polar. Organic molecules are sorted into classes according to the principal functional group(s) each contains. Using a separation mode based on polarity, the relative chromatographic retention of different kinds of molecules is largely determined by the nature and location of these functional groups. As shown in Figure P, classes of molecules can be ordered by their relative retention into a range or spectrum of chromatographic polarity from highly polar to highly non-polar.
Figure P: Chromatographic Polarity Spectrum by Analyte Functional Group
Water [a small molecule with a high dipole moment] is a polar compound. Benzene [an aromatic hydrocarbon] is a non-polar compound. Molecules with similar chromatographic polarity tend to be attracted to each other; those with dissimilar polarity exhibit much weaker attraction, if any, and may even repel one another. This becomes the basis for chromatographic separation modes based on polarity.
Another way to think of this is by the familiar analogy: oil [non-polar] and water [polar] don’t mix. Unlike in magnetism where opposite poles attract each other, chromatographic separations based on polarity depend upon the stronger attraction between likes and the weaker attraction between opposites. Remember, “Like attracts like” in polarity-based chromatography.


Figure Q: Proper Combination of Mobile and Stationary Phases Effects Separation Based on Polarity

To design a chromatographic separation system [see Figure Q], we create competition for the various compounds contained in the sample by choosing a mobile phase and a stationary phase with different polarities. Then, compounds in the sample that are similar in polarity to the stationary phase [column packing material] will be delayed because they are more strongly attracted to the particles. Compounds whose polarity is similar to that of the mobile phase will be preferentially attracted to it and move faster.
In this way, based upon differences in the relative attraction of each compound for each phase, a separation is created by changing the speeds of the analytes.
Figures R-1, R-2, and R-3 display typical chromatographic polarity ranges for mobile phases, stationary phases, and sample analytes, respectively. Let’s consider each in turn to see how a chromatographer chooses the appropriate phases to develop the attraction competition needed to achieve a polarity-based HPLC separation.


Figure R-1: Mobile Phase Chromatographic Polarity Spectrum
A scale, such as that shown in Figure R-1, upon which some common solvents are placed in order of relative chromatographic polarity is called an eluotropic series. Mobile phase molecules that compete effectively with analyte molecules for the attractive stationary phase sites displace these analytes, causing them to move faster through the column [weakly retained]. Water is at the polar end of mobile-phase-solvent scale, while hexane, an aliphatic hydrocarbon, is at the non-polar end. In between, single solvents, as well as miscible-solvent mixtures [blended in proportions appropriate to meet specific separation requirements], can be placed in order of elution strength. Which end of the scale represents the ‘strongest’ mobile phase depends upon the nature of the stationary phase surface where the competition for the analyte molecules occurs.
Figure R-2: Stationary Phase Particle Chromatographic Polarity Spectrum

Silica has an active, hydrophilic [water-loving] surface containing acidic silanol [silicon-containing analog of alcohol] functional groups. Consequently, it falls at the polar end of the stationary-phase scale shown in Figure R-2. The activity or polarity of the silica surface may be modified selectively by chemically bonding to it less polar functional groups [bonded phase]. Examples shown here include, in order of decreasing polarity, cyanopropylsilyl- [CN], n-octylsilyl- [C8], and n-octadecylsilyl- [C18, ODS] moieties on silica. The latter is a hydrophobic [water-hating], very non-polar packing.




Figure R-3: Compound/Analyte Chromatographic Polarity Spectrum


Figure R-3 repeats the chromatographic polarity spectrum of our sample [shown in Figure P]. After considering the polarity of both phases, then, for a given stationary phase, a chromatographer must choose a mobile phase in which the analytes of interest are retained, but not so strongly that they cannot be eluted. Among solvents of similar strength, the chromatographer considers which phase combination may best exploit the more subtle differences in analyte polarity and solubility to maximize the selectivity of the chromatographic system. Like attracts like, but, as you probably can imagine from the discussion so far, creating a separation based upon polarity involves knowledge of the sample and experience with various kinds of analytes and retention modes. To summarize, the chromatographer will choose the best combination of a mobile phase and particle stationary phase with appropriately opposite polarities. Then, as the sample analytes move through the column, the rule like attracts like will determine which analytes slow down and which proceed at a faster speed.
Normal-Phase HPLC
In his separations of plant extracts, Tswett was successful using a polar stationary phase [chalk in a glass column; see Figure A] with a much less polar [non-polar] mobile phase. This classical mode of chromatography became known as normal phase.

Figure S-1: Normal-Phase Chromatography
Figure S-1 represents a normal-phase chromatographic separation of our three-dye test mixture. The stationary phase is polar and retains the polar yellow dye most strongly. The relatively non-polar blue dye is won in the retention competition by the mobile phase, a non-polar solvent, and elutes quickly. Since the blue dye is most like the mobile phase [both are non-polar], it moves faster. It is typical for normal-phase chromatography on silica that the mobile phase is 100% organic; no water is used.

Reversed-Phase HPLC
The term reversed-phase describes the chromatography mode that is just the opposite of normal phase, namely the use of a polar mobile phase and a non-polar [hydrophobic] stationary phase. Figure S-2 illustrates the black three-dye mixture being separated using such a protocol.

Figure S-2: Reversed-Phase Chromatography
Now the most strongly retained compound is the more non-polar blue dye, as its attraction to the non-polar stationary phase is greatest. The polar yellow dye, being weakly retained, is won in competition by the polar, aqueous mobile phase, moves the fastest through the bed, and elutes earliest like attracts like.
Today, because it is more reproducible and has broad applicability, reversed-phase chromatography is used for approximately 75% of all HPLC methods. Most of these protocols use as the mobile phase an aqueous blend of water with a miscible, polar organic solvent, such as acetonitrile or methanol. This typically ensures the proper interaction of analytes with the non-polar, hydrophobic particle surface. A C18–bonded silica [sometimes called ODS] is the most popular type of reversed-phase HPLC packing.
Table C presents a summary of the phase characteristics for the two principal HPLC separation modes based upon polarity. Remember, for these polarity-based modes, like attracts like.


Table C: Phase Characteristics for Separations Based on Polarity

Hydrophilic-Interaction Chromatography [HILIC]
HILIC may be viewed as a variant of normal-phase chromatography. In normal-phase chromatography, the mobile phase is 100% organic. Only traces of water are present in the mobile phase and in the pores of the polar packing particles. Polar analytes bind strongly to the polar stationary phase and may not elute.

Adding some water [< 20%] to the organic mobile phase [typically an aprotic solvent like acetonitrile] makes it possible to separate and elute polar compounds that are strongly retained in the normal-phase mode [or weakly retained in the reversed-phase mode]. Water, a very polar solvent, competes effectively with polar analytes for the stationary phase. HILIC may be run in either isocratic or gradient elution modes. Polar compounds that are initially attracted to the polar packing material particles can be eluted as the polarity [strength] of the mobile phase is increased [by adding more water]. Analytes are eluted in order of increasing hydrophilicity [chromatographic polarity relative to water]. Buffers or salts may be added to the mobile phase to keep ionizable analytes in a single form.

Hydrophobic-Interaction Chromatography [HIC]
HIC is a type of reversed-phase chromatography that is used to separate large biomolecules, such as proteins. It is usually desirable to maintain these molecules intact in an aqueous solution, avoiding contact with organic solvents or surfaces that might denature them. HIC takes advantage of the hydrophobic interaction of large molecules with a moderately hydrophobic stationary phase, e.g., butyl-bonded [C4], rather than octadecyl-bonded [C18], silica. Initially, higher salt concentrations in water will encourage the proteins to be retained [salted out] on the packing. Gradient separations are typically run by decreasing salt concentration. In this way, biomolecules are eluted in order of increasing hydrophobicity.

Separations Based on Charge: Ion-Exchange Chromatography [IEC]
For separations based on polarity, like is attracted to like and opposites may be repelled. In ion-exchange chromatography and other separations based upon electrical charge, the rule is reversed. Likes may repel, while opposites are attracted to each other. Stationary phases for ion-exchange separations are characterized by the nature and strength of the acidic or basic functions on their surfaces and the types of ions that they attract and retain. Cation exchange is used to retain and separate positively charged ions on a negative surface. Conversely, anion exchange is used to retain and separate negatively charged ions on a positive surface [see Figure T]. With each type of ion exchange, there are at least two general approaches for separation and elution.

Figure T: Ion-Exchange Chromatography

Strong ion exchangers bear functional groups [e.g., quaternary amines or sulfonic acids] that are always ionized. They are typically used to retain and separate weak ions. These weak ions may be eluted by displacement with a mobile phase containing ions that are more strongly attracted to the stationary phase sites. Alternately, weak ions may be retained on the column, then neutralized by in situ changing the pH of the mobile phase, causing them to lose their attraction and elute.
Weak ion exchangers [e.g., with secondary-amine or carboxylic-acid functions] may be neutralized above or below a certain pH value and lose their ability to retain ions by charge. When charged, they are used to retain and separate strong ions. If these ions cannot be eluted by displacement, then the stationary phase exchange sites may be neutralized, shutting off the ionic attraction, and permitting elution of the charged analytes.


Table D: Ion-Exchange Guidelines

When weak ion exchangers are neutralized, they may retain and separate species by hydrophobic [reversed-phase] or hydrophilic [normal-phase] interactions; in these cases, elution strength is determined by the polarity of the mobile phase [Figure R-1]. Thus, weak ion exchangers may be used for mixed-mode separations [separations based on both polarity and charge].
Table D outlines guidelines for the principal categories of ion exchange. For example, to retain a strongly basic analyte [always positively charged], use a weak-cation-exchange stationary phase particle at pH > 7; this assures a negatively charged particle surface. To release or elute the strong base, lower the pH of the mobile phase below 3; this removes the surface charge and shuts off the ion-exchange retention mechanism.
Note that a pKa is the pH value at which 50% of the functional group is ionized and 50% is neutral. To assure an essentially neutral, or a fully charged, analyte or particle surface, the pH must be adjusted to a value at least 2 units beyond the pKa, as appropriate [indicated in Table D].
Do not use a strong-cation exchanger to retain a strong base; both remain charged and strongly attracted to each other, making the base nearly impossible to elute. It can only be removed by swamping the strong cation exchanger with a competing base that exhibits even stronger retention and displaces the compound of interest by winning the competition for the active exchange sites. This approach is rarely practical, or safe, in HPLC and SPE. [Very strong acids and bases are dangerous to work with, and they may be corrosive to materials of construction used in HPLC fluidics!]
Separations Based on Size: Size-Exclusion Chromatography [SEC] –
Gel-Permeation Chromatography [GPC]
In the 1950s, Porath and Flodin discovered that biomolecules could be separated based on their size, rather than on their charge or polarity, by passing, or filtering, them through a controlled-porosity, hydrophilic dextran polymer. This process was termed gel filtration. Later, an analogous scheme was used to separate synthetic oligomers and polymers using organic-polymer packings with specific pore-size ranges. This process was called gel-permeation chromatography [GPC]. Similar separations done using controlled-porosity silica packings were called size-exclusion chromatography [SEC]. Introduced in 1963, the first commercial HPLC instruments were designed for GPC applications [see Reference 3].

All of these techniques are typically done on stationary phases that have been synthesized with a pore-size distribution over a range that permits the analytes of interest to enter, or to be excluded from, more or less of the pore volume of the packing. Smaller molecules penetrate more of the pores on their passage through the bed. Larger molecules may only penetrate pores above a certain size so they spend less time in the bed. The biggest molecules may be totally excluded from pores and pass only between the particles, eluting very quickly in a small volume. Mobile phases are chosen for two reasons: first, they are good solvents for the analytes; and, second, they may prevent any interactions [based on polarity or charge] between the analytes and the stationary phase surface. In this way, the larger molecules elute first, while the smaller molecules travel slower [because they move into and out of more of the pores] and elute later, in decreasing order of their size in solution. Hence the simple rule: Big ones come out first.

Since it is possible to correlate the molecular weight of a polymer with its size in solution, GPC revolutionized measurement of the molecular-weight distribution of polymers that, in turn, determines the physical characteristics that may enhance, or detract from, polymer processing, quality, and performance [how to tell good from bad polymer].

Conclusion
We hope you have enjoyed this brief introduction to HPLC. We encourage you to read the references below and to study the Appendix on HPLC Nomenclature.

HPLC Column Hardware

A column tube and fittings must contain the chromatographic packing material [stationary phase] that is used to effect a separation. It must withstand backpressure created both during manufacture and in use. Also, it must provide a well-controlled [leak-free, minimum-volume, and zero-dead-volume] flow path for the sample at its inlet, and analyte bands at its outlet, and be chemically inert relative to the separation system [sample, mobile, and stationary phases]. Most columns are constructed of stainless steel for highest pressure resistance. PEEK[an engineered plastic] and glass, while less pressure tolerant, may be used when inert surfaces are required for special chemical or biological applications. [Figure M-1].

Figure M-1: Column Hardware Examples

A glass column wall offers a visual advantage. In the photo in Figure M-2, flow has been stopped while the sample bands are still in the column. You can see that the three dyes in the injected sample mixture have already separated in the bed; the yellow analyte, traveling fastest, is just about to exit the column.


Figure M-2: A Look Inside a Column
Separation Performance – Resolution
The degree to which two compounds are separated is called chromatographic resolution [RS]. Two principal factors that determine the overall separation power or resolution that can be achieved by an HPLC column are: mechanical separation power, created by the column length, particle size, and packed-bed uniformity, and chemical separation power, created by the physicochemical competition for compounds between the packing material and the mobile phase. Efficiency is a measure of mechanical separation power, while selectivity is a measure of chemical separation power.
Mechanical Separation Power – Efficiency
If a column bed is stable and uniformly packed, its mechanical separation power is determined by the column length and the particle size. Mechanical separation power, also called efficiency, is often measured and compared by a plate number [symbol = N]. Smaller-particle chromatographic beds have higher efficiency and higher backpressure. For a given particle size, more mechanical separation power is gained by increasing column length. However, the trade-offs are longer chromatographic run times, greater solvent consumption, and higher backpressure. Shorter column lengths minimize all these variables but also reduce mechanical separation power, as shown in Figure N.

Figure N: Column Length and Mechanical Separating Power [Same Particle Size]
Figure O: Particle Size and Mechanical Separating Power [Same Column Length]
For a given particle chemistry, mobile phase, and flow rate, as shown in Figure O, a column of the same length and i.d., but with a smaller particle size, will deliver more mechanical separation power in the same time. However, its backpressure will be much higher.
Chemical Separation Power – Selectivity
The choice of a combination of particle chemistry [stationary phase] and mobile-phase composition—the separation system—will determine the degree of chemical separation power [how we change the speed of each analyte]. Optimizing selectivity is the most powerful means of creating a separation; this may obviate the need for the brute force of the highest possible mechanical efficiency. To create a separation of any two specified compounds, a scientist may choose among a multiplicity of phase combinations [stationary phase and mobile phase] and retention mechanisms [modes of chromatography]. These are discussed in the next section.

Identifying and Quantitating Compounds

In Figure H, three dye compounds are represented by three peaks separated in time in the chromatogram. Each elutes at a specific location, measured by the elapsed time between the moment of injection [time zero] and the time when the peak maximum elutes. By comparing each peak’s retention time [tR] with that of injected reference standards in the same chromatographic system [same mobile and stationary phase], a chromatographer may be able to identify each compound.

Figure I-1: Identification

In the chromatogram shown in Figure I-1, the chromatographer knew that, under these LC system conditions, the analyte, acrylamide, would be separated and elute from the column at 2.85 minutes [retention time]. Whenever a new sample, which happened to contain acrylamide, was injected into the LC system under the same conditions, a peak would be present at 2.85 minutes [see Sample B in Figure I-2].
[For a better understanding of why some compounds move more slowly [are better retained] than others, please review the HPLC Separation Modes section on page 28].
Once identity is established, the next piece of important information is how much of each compound was present in the sample. The chromatogram and the related data from the detector help us calculate the concentration of each compound. The detector basically responds to the concentration of the compound band as it passes through the flow cell. The more concentrated it is, the stronger the signal; this is seen as a greater peak height above the baseline.

Figure I-2: Identification and Quantitation

In Figure I-2, chromatograms for Samples A and B, on the same time scale, are stacked one above the other. The same volume of sample was injected in both runs. Both chromatograms display a peak at a retention time [tR] of 2.85 minutes, indicating that each sample contains acrylamide. However, Sample A displays a much bigger peak for acrylamide. The area under a peak [peak area count] is a measure of the concentration of the compound it represents. This area value is integrated and calculated automatically by the computer data station. In this example, the peak for acrylamide in Sample A has 10 times the area of that for Sample B. Using reference standards, it can be determined that Sample A contains 10 picograms of acrylamide, which is ten times the amount in Sample B [1 picogram]. Note there is another peak [not identified] that elutes at 1.8 minutes in both samples. Since the area counts for this peak in both samples are about the same, this unknown compound may have the same concentration in both samples.
Isocratic and Gradient LC System Operation
Two basic elution modes are used in HPLC. The first is called isocratic elution. In this mode, the mobile phase, either a pure solvent or a mixture, remains the same throughout the run. A typical system is outlined in Figure J-1.


Figure J-1: Isocratic LC System

The second type is called gradient elution, wherein, as its name implies, the mobile phase composition changes during the separation. This mode is useful for samples that contain compounds that span a wide range of chromatographic polarity [see section on HPLC Separation Modes]. As the separation proceeds, the elution strength of the mobile phase is increased to elute the more strongly retained sample components.

Figure J-2: High-Pressure-Gradient System

In the simplest case, shown in Figure J-2, there are two bottles of solvents and two pumps. The speed of each pump is managed by the gradient controller to deliver more or less of each solvent over the course of the separation. The two streams are combined in the mixer to create the actual mobile phase composition that is delivered to the column over time. At the beginning, the mobile phase contains a higher proportion of the weaker solvent [Solvent A]. Over time, the proportion of the stronger solvent [Solvent B] is increased, according to a predetermined timetable. Note that in Figure J-2, the mixer is downstream of the pumps; thus the gradient is created under high pressure. Other HPLC systems are designed to mix multiple streams of solvents under low pressure, ahead of a single pump. A gradient proportioning valve selects from the four solvent bottles, changing the strength of the mobile phase over time [see Figure J-3].


Figure J-3: Low-Pressure-Gradient System


HPLC Scale [Analytical, Preparative, and Process]
We have discussed how HPLC provides analytical data that can be used both to identify and to quantify compounds present in a sample. However, HPLC can also be used to purify and collect desired amounts of each compound, using a fraction collector downstream of the detector flow cell. This process is called preparative chromatography [see Figure K].
In preparative chromatography, the scientist is able to collect the individual analytes as they elute from the column [e.g., in this example: yellow, then red, then blue].

Figure K: HPLC System for Purification: Preparative Chromatography
The fraction collector selectively collects the eluate that now contains a purified analyte, for a specified length of time. The vessels are moved so that each collects only a single analyte peak.
A scientist determines goals for purity level and amount. Coupled with knowledge of the complexity of the sample and the nature and concentration of the desired analytes relative to that of the matrix constituents, these goals, in turn, determine the amount of sample that needs to be processed and the required capacity of the HPLC system. In general, as the sample size increases, the size of the HPLC column will become larger and the pump will need higher volume-flow-rate capacity. Determining the capacity of an HPLC system is called selecting the HPLC scale. Table A lists various HPLC scales and their chromatographic objectives.


Table A: Chromatography Scale


The ability to maximize selectivity with a specific combination of HPLC stationary and mobile phases—achieving the l the requirements for scaling up a separation [see discussion on HPLC Separation Modes]. Capacity then becomes a matter of scaling the column volume [Vc] to the amount of sample to be injected and choosing an appropriate particle size [determines pressure and efficiency; see discussion of Separation Power]. Column volume, a function of bed length [L] and internal diameter [i.d.], determines the amount of packing material [particles] that can be contained (see Figure L).

Figure L: HPLC Column Dimensions

In general, HPLC columns range from 20 mm to 500 mm in length [L] and 1 mm to 100 mm in internal diameter [i.d.]. As the scale of chromatography increases, so do column dimensions, especially the cross-sectional area. To optimize throughput, mobile phase flow rates must increase in proportion to cross-sectional area. If a smaller particle size is desirable for more separation power, pumps must then be designed to sustain higher mobile-phase-volume flow rates at high backpressure. Table B presents some simple guidelines on selecting the column i.d. and particle size range recommended for each scale of chromatography.
For example, a semi-preparative-scale application [red X] would use a column with an internal diameter of 10–40 mm containing 5–15 micron particles. Column length could then be calculated based on how much purified compound needs to be processed during each run and on how much separation power is required.


Table B: Chromatography Scale vs. Column Diameter and Particle Size